ACTA Pharmaceutica Sciencia 2011 , Vol 53 , Num 4
THE COMPARISON OF THE OIL POLLUTION RESULTS IN THE SEA WATER BY TWO DIFFERENT UVF APPARATUS
In this work was discussed the results of obtained by two different apparatus as Shimadzu RF 1501 and Jasco FP 6300. The measurements of oil pollution level in the sea waters were determined by using references of the Russian and Iraq crude oils and chrysene. It was found some differences between the results obtained by two apparatus. The error in the results of the references are in Russian crude oil between 01-4.4%, in Iraq crude oil 2% and in chrysene sample 5.8%. Based on these results it was suggested that the measurements must be made by the same apparatus. Keywords : UVF, APPARATUS, SHIMADZU, JASCO